Siemens SIPROCESS GA700 Compact Operating Instructions page 187

Continuous gas analysis, devices in explosion-proof design
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

Table 8-3
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Service interface (rear)
Service interface (front)
Table 8-4
Option modules
Option module 1.1
Option module 2.1
Option module 2.2
Table 8-5
Climatic conditions
Permissible operating height
Ambient temperature and humidity
Devices in explosion-proof design
Compact Operating Instructions, 05/2015, A5E35134047-02
Rack-mounted enclosure: Electrical inputs and outputs
Rack-mounted enclosure: Option modules
Rack-mounted enclosure: Climatic conditions
8.2 General technical specifications
8, with changeover contacts, parameters can be freely as‐
signed, e.g. for measuring range identification; max. load: 24
V AC/DC/1.7 A (total load for all 8 relay outputs in continuous
operation max. 160 W), floating, non-sparking
8, designed for 24 V, isolated, freely parameterizable, e.g. for
autoranging
Ethernet RJ 45, 100 MBit
Ethernet RJ 45, 100 MBit
● 12 digital outputs, with changeover contacts, max. load:
24 V AC/DC / 1.7
(total load for all 12 relay outputs in continuous operation
max. 244 W), floating, non-sparking
● 8 digital inputs, designed for 24 V, floating, freely
configurable
● 6 analog outputs, 0/4 to 20 mA, floating; maximum load:
R
≤ 750 Ω
L
● 6 digital outputs, load rating: 24 V AC/DC/1.7 A (total load
for all 6 relay outputs in continuous operation max. 122
W), floating, non-sparking
● 6 analog outputs, 0/4 to 20 mA, floating; maximum load:
R
≤ 750 Ω
L
● 4 analog inputs 0/4 to 20 mA, non-isolated, internal
resistance ≤ 100 Ω
● 4 digital inputs, designed for 24 V, floating
3 000 m above sea level sea level
→ Technical specifications for operation in hazardous areas
(Page 88)
Technical specifications
81

Advertisement

Chapters

Table of Contents
loading

Table of Contents