Nonin Avant 9600 Operator's Manual page 42

Digital pulse oximeter
Hide thumbs Also See for Avant 9600:
Table of Contents

Advertisement

T r o u b l e s h o o t i n g
Problem
A dash (-) appears in
the %SpO
display.
2
An error code appears
in the display area.
39
Possible Cause
A poor signal from the
finger is being detected.
The finger was removed
from the sensor.
The Avant 9600 is not
functioning.
The Avant 9600
encountered an error.
Possible Solution
Reposition the finger or insert
a different finger and keep the
sensor motionless for at least
10 seconds.
Position the sensor at a
different site.
Reinsert the finger and keep
the sensor motionless for at
least 10 seconds.
Turn the unit off, check all
connections, and retry.
Contact NONIN Customer
Support.
1. Turn the unit off and then
back on again to remove
the error code.
2. If the error persists,
disconnect all power (AC
and battery), and then
reconnect the power and
turn the unit back on.
3. If the error still persists,
note the error code and
contact NONIN
Customer Support.