Problem
The unit is in Alarm
mode, but no audible
alarms can be heard.
The Avant 9600 does
not record data.
If these solutions do not correct the problem, please contact NONIN Customer
Support at (800) 356-8874 (USA and Canada) or +1 (763) 553-9968.
Possible Cause
The 2-minute Alarm
Silence button is
activated.
Option switch 1 is in the
UP position, and the
unit's volume is set to
zero.
The battery is low.
The battery is missing.
Possible Solution
Press the Alarm Silence button
to re-engage alarm volume, or
wait for two minutes—and
alarm tones will automatically
re-engage.
Adjust the alarm volume, or
return option switch 1 to the
DOWN position if you desire
audible alarms.
Recharge the battery.
Contact your distributor or
NONIN Customer Support
for repair or replacement.
40