Serial Patient Data Outputs - Nonin Avant 9600 Operator's Manual

Digital pulse oximeter
Hide thumbs Also See for Avant 9600:
Table of Contents

Advertisement

Serial Patient Data Outputs

The Avant 9600 features real-time and print-on-demand capabilities. All outputs
include a header containing model number, time, and date information.
The Avant 9600 provides real-time data output capability via the RS232 connector
port. A null modem cable may be used to connect the Avant 9600 to the receiving
computer. The information from the Avant 9600 is sent in an ASCII serial format
at 9600 baud with 8 data bits, 1 start bit, and 2 stop bits. Each line is terminated by
CR/LF.
Print-On-Demand Output (Default)
The Avant 9600 features a Print-On-Demand option, allowing users to output data
each time the ON/STANDBY button is pressed.
Real-Time Patient Data Output
Data are sent once per second by the Avant 9600 in one of the following formats:
• If option switch 8 is in the UP position, the data will be displayed as follows:
where XXX and YYY are the fast-responding SpO
• If option switch 8 is in the DOWN position, the data will be displayed
as follows:
where XXX and YYY are the SpO
option switch 2.
NOTE: Pressing the ON/STANDBY button will insert a "*" at the end of
the corresponding printed line to serve as an event marker.
SPO2=XXX HR=YYY F
SPO2=XXX HR=YYY
and pulse rate values as selected by
2
and pulse rate values.
2
28