Patient Alarms
Only one alarm condition can be reported to the VueLink at a time. Because of this,
patient alarm conditions are assigned priorities—with only the highest priority
alarm displayed at any one time. Alarms are sent to the VueLink even when the
oximeter's alarms are silenced. Alarm priorities are assigned as follows:
1. SpO
2. Pulse Rate High Limit
3. Pulse Rate Low Limit
4. SpO
5. Low Perfusion
Equipment Alarms
VueLink displays equipment alarms as "inoperatives." Only one inoperative
condition can be reported to the VueLink at a time. Because of this, inoperatives are
assigned priorities—with only the highest priority displayed at any one time.
Inoperatives are sent to the VueLink even when the oximeter's alarms are silenced.
Inoperative priorities are assigned as follows:
1. No SpO
2. SpO
3. No SpO
4. No Pulse Data
5. Low Battery
The following text strings, associated with the Avant 9600, are sent through the
VueLink module for display on the multiparameter monitor:
Parameter
Displayed High SpO2 Limit
Displayed Low SpO2 Limit
Displayed High Pulse Rate Limit
Displayed Low Pulse Rate Limit
SpO2 High Limit Alarm
SpO2 Low Limit Alarm
Pulse Rate High Limit Alarm
Pulse Rate Low Limit Alarm:
Low Perfusion Alarm
Sensor Disconnect
Sensor Fault
No SpO2 Data Available
No Pulse Data Available
Low Battery
Low Limit
2
High Limit
2
Sensor
2
Sensor Fault
2
Data
2
Displayed Text
HiSpO2
LoSpO2
Hi PR
Lo PR
SpO2: HIGH
SpO2: LOW
PULSE: HIGH
PULSE: LOW
LOW PERFUSION
NO SpO2 SENSOR
SpO2 SENSOR
NO SpO2 Data
NO Pulse Data
Battery: Pulse Ox
30