Avant 9600 Option switches

The Avant 9600 contains eight option switches that are located under the blue back
cover of the unit (See "Installing the Batteries" for instructions on removing the
back cover). The UP position faces the top of the unit, and the DOWN position
faces the bottom of the unit. The factory setting for all Avant 9600 option
switches is the DOWN position.
Switch
Function
Switch 1
Alarm Disable Lock
Up—Alarm volume may be disabled
Down—Alarm volume cannot be set to zero
Switch 2
Normal / Slow SpO
Up—Slow Averaging (8 beat exponential average)
Down—Normal Averaging (4 beat exponential average)
Switch 3
Alarm Unlatched / Latched
Up—Alarms Latched
Down—Alarms Unlatched
Switch 4
Factory / User-Defined Defaults
Up—User-Defined Defaults for Alarm Limits and Volume Settings
Down—Factory Defaults for Alarm Limits and Volume Settings
US / International Date Format
Switch 5
Up—International Date Format
Down—US Date Format
Switch 6
Power Saving Display Dim During Battery Operation
Up—Power Saving Disabled
Down—Power Saving Enabled
Nurse Call Output
Switch 7
Up—Continuous
Down—Momentary
Switch 8
Serial Patient Data Output
Up—Fast responding SpO
Down—SpO
Averaging
2
and pulse rate output
2
and pulse rate output, as selected by option switch 2