Nonin Onyx 9500 Instructions For Use

Nonin Onyx 9500 Instructions For Use

Finger pulse oximeter

Advertisement

Quick Links

Instructions for Use

®
Onyx
9500 Finger Pulse Oximeter
0123

Indications for Use

®
®
The NONIN
Onyx
Model 9500 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in measuring
and displaying functional oxygen saturation of arterial hemoglobin (%SpO
perfused. It is intended for spot-checking of adult and pediatric patients on fingers (other than the thumb) between 0.3 - 1.0 inch
(0.8 – 2.5 cm) thick.
Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.
Contraindications
Do not use this device in an MRI environment, in an explosive atmosphere, or on infant or neonatal patients.
This device is not defibrillation proof per IEC 60601-1:1990 clause 17h.
Warnings
This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
The device must be able to measure the pulse properly to obtain an accurate SpO
hindering the pulse measurement before relying on the SpO
Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
General operation of the device may be affected by the use of an electrosurgical unit (ESU).
The use of accessories other than those specified in these instructions may result in increased electromagnetic emission
and/or decreased immunity of this device.
This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the
device should be observed carefully to verify normal operation.
Cautions
This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Significant levels
of dysfunctional hemoglobin may affect measurement accuracy.
This device has no audible alarms and is intended only for spot-checking.
Inspect the sensor application site at least every 6 to 8 hours to ensure correct sensor alignment and skin integrity. Patient
sensitivity to sensors may vary due to medical status or skin condition.
Fluctuating or very bright light, moisture, blood pressure cuffs, infusion lines, venous pulsations, insufficient pulse signals,
anemia, arterial catheters, nail polish, and/or artificial nails may degrade the device's performance.
The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
This device's display will go blank after 30 seconds of no readings or poor readings.
In some circumstances, the device may interpret motion as good pulse quality. Minimize patient motion as much as possible.
Cardiogreen and other intravascular dyes may affect the accuracy of the SpO
Do not autoclave or immerse this device in liquid or use caustic or abrasive cleaning agents.
A flexible circuit connects the two halves. Do not twist or pull the flexible circuit or overextend the device's spring.
Do not hang the lanyard from the device's flexible circuit/strain relief.
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.
This equipment complies with IEC 60601-1-2:2001 for electromagnetic compatibility for medical electrical equipment and/
or systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical
installation. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical
noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or
strength of a source might disrupt the performance of this device. Medical electrical equipment needs special precautions
regarding EMC, and all equipment must be installed and put into service according to the EMC information specified in this
manual.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than
30 days. Do not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries
at the same time. These actions may cause the batteries to leak.
Follow local, state and national governing ordinances and recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not
dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor
regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call NONIN
distributor's contact information.
Symbol
!
0123
SN
U L
C
US
EC
REP
IPX2
) and pulse rate of patients who are well or poorly
2
measurement.
2
Symbols
Definition of Symbol
Consult Instructions for Use
Caution
CE Marking indicating conformance to EC Directive No. 93/42/EEC concerning medical
devices
Type BF Applied Part (patient isolation from electrical shock)
Not for Continuous Monitoring (no alarm for SpO
Serial Number
Battery Orientation
UL Mark for Canada and the United States with respect to electric shock, fire, and mechanical
hazards only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1
Lot Number
Indicates separate collection for electrical and electronic equipment (WEEE).
Non-ionizing electromagnetic radiation. Equipment includes RF transmitters.
Interference may occur in the vicinity of equipment marked with this symbol.
Authorized Representative in the European Community.
Protected against vertically falling water drops when enclosure is tilted up to 15 degrees,
per IEC 60529.
measurement. Verify that nothing is
2
measurement.
2
®
for your
)
2

Advertisement

loading

Summary of Contents for Nonin Onyx 9500

  • Page 1: Instructions For Use

    This device contains WEEE materials; please contact your distributor ® regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call NONIN for your distributor’s contact information.
  • Page 2: Installing Batteries

    ® ® Two 1.5 volt AAA-size batteries power the Onyx for about 1,600 spot checks or 18 hours of continuous operation. NONIN ® recommends using alkaline batteries (included with each new Onyx ). When batteries are low, the numeric displays flash once per second.
  • Page 3: Testing Summary

    ® ® ® Onyx , shall void the warranty. All non-warranty work shall be done at NONIN ’s standard rates and charges in effect at the time of delivery to NONIN...
  • Page 4: Manufacturer's Declaration

    Manufacturer’s Declaration Refer to the following tables for specific information regarding this device’s compliance to IEC 60601-1-2:2001. Table 1: Electromagnetic Emissions Emissions Test Compliance Electromagnetic Environment—Guidance This device is intended for use in the electromagnetic environment specified below. The customer and/or user of this device should ensure that it is used in such an environment. RF Emissions Group 1 This device uses RF energy only for its internal function.
  • Page 5 NONIN Medical, Inc. 0123 • 800-356-8874 (US & Canada) 13700 1st Avenue North Authorized EC Representative: • Fax +1 (763) 553-7807 Plymouth, MN 55441-5443 USA MPS, Medical Product Service • E-mail: mail@nonin.com © 2006 GmbH • www.nonin.com 4821-001-06 Borngasse 20...