Default Settings

The Avant 9700 features Factory Default and User-Defined Default settings.
Factory Default Setting
In Factory Default setting, all adjustable alarm and volume parameters are set at
their default values. Factory Default setting is the Avant 9700's default operating
setting. It is indicated by DIP switch 4 in the DOWN position.
User-Defined Default Setting
In User-Defined Default Setting (DIP switch 4 in the UP position), alarm limit and
volume settings must be adjusted. When this setting is first activated, valid limit
settings must be entered for SpO
alarm limits, pulse rate alarm limits, alarm
2
volume, and pulse rate volume; the Avant 9700 will not return to normal operating
mode until all default values have been set. Once set, the adjusted values are used as
defaults until the Avant 9700 is turned on with DIP switch 4 in the DOWN position,
at which time the device restarts in Factory Default setting.
NOTE: Patient Security mode overrides User-Defined Default setting.
NOTE: In User-Defined Default setting, the lower SpO
alarm limit
2
cannot be set below 80.
11