SYSTEM
Temperature
Operating
Storage/Transportation
Humidity
Operating
Storage/Transportation
Power Requirements
Battery Life
Operating
Storage
Recharge
Dimensions
Weight
Memory
Classifications per IEC 60601- 1 / CSA601.1 / UL2601-1
Type of Protection
Degree of Protection
Enclosure Degree of
Ingress Protection
Mode of Operation
+32° to +122°F (+0° to +50°C)
-22° to +122°F (-30° to +50°C)
10% to 90% noncondensing
10% to 95% noncondensing
7.2 volt battery pack (6 cells)
or 12 VDC 1.5A AC Adapter
minimum 12 hours of continuous operation
with a fully charged battery pack
27 days
4 hours
5.5" H x 7.25" W x 4.5" D
2.2 lbs
115 hours minimum
Class I (when on AC power with 300PS-
UNIV battery charger)
Internally powered (on battery
power)
Type BF-Applied Part
IPX0
Continuous
32