Understanding cabinet LEDs

Figure 55 Analytical unit status LEDs
On the analytical unit cabinet, only one LED at a time can light up, as such:
Solid
RED
An alarm threshold has been reached by any of
the measured gases. For more information on
gas alarm thresholds, refer to the CoreSense
M10 User Guide.
ALARM
YELLOW
A warning threshold has been reached by any of
the measured gases. For more information on
gas warning thresholds, refer to the CoreSense
M10 User Guide.
WARNING
GREEN
Normal operating conditions.
NORMAL
BLUE
User attention is required as the sensor is either
operating outside its nominal specifications
or experimenting a fault condition. Consult the
event log and contact ABB.
MAINTENANCE
72
Installation Guide
Blinking
N/A
The operating system is booting up.
The cabinet is starting up.
User attention is required as the last working
pump or source is being used. An event is
created accordingly and a service call should be
made.
NOTE: The blinking blue LED can be lit at
the same time as other LEDs since the
unresponsive source/pump problem can
happen independently of other problems.