Installing firmware updates

At some point in the future, you might be asked by ABB service representatives to update your system
firmware. This can be done with a web browser (remotely or locally) or a USB key (locally).
Before performing any firmware update, it is strongly recommended to export your system logs and
history files (refer to the LWT300 User Guide for more information on these procedures).

Updating with a web browser

To update the firmware via a web browser:
1 Skip to step 2 if you are working remotely. Otherwise, open the CoreSense M10 cabinet and connect
your laptop to an Ethernet communication port (SCADA or SERVICE) of the analytical unit with a
straight RJ45 Ethernet cable.
2 Open your web browser and point it to the required IP address:
– if connected to the SCADA port: the static IP address (
indicated on the local HMI;
– if connected to the SERVICE port:
This address points to the CoreSense M10 dashboard.
3 From the CoreSense M10 dashboard, click Settings. The Settings page appears.
4 Click Update firmware and enter your password (if necessary). The firmware update page appears.
Figure 60 Firmware update page
5 Click Choose File.
6 In the Open window that appears, locate and select the firmware file (.zip) sent to you by the ABB
service representative.
7 Click Open. You return to the firmware update page. The name of the file that you selected appears
next to the Choose File button and the Update button is now active.
8 Click Update.
The file is uploaded to your system and the firmware update takes place.
NOTICE
Do not exit the browser during the firmware update process.
76