Placing The Unit On The Shock Absorbers - ABB CoreSense M10 Installation Manual

Multi-gas monitoring system
Hide thumbs Also See for CoreSense M10:
Table of Contents

Advertisement

Placing the unit on the shock absorbers

WARNING
The CoreSense M10 analytical unit weighs over 65 kg (144 lb). It could cause serious
injuries or even death if it were to tip over and/or fall on someone.
Once the mounting posts and shock absorbers are properly installed, you need to position the
analytical unit.
ABB recommends using a sling and a crane to lift the unit up on to the mounting posts. To this end,
slots are designed in the unit mounting frame to help lift up the unit.
Figure 18 Lifting slots on back of unit
However, the following procedure explains how an installation team should install the unit by hand. It is
important to follow the general lifting and handling precautions given on page 7.
To place the unit on the mounting posts:
1 With one team member on each side of the analytical unit, lift the unit upright.
2 With proper lifting technique, lift up the analytical unit and lower it on the four shock absorbers.
Make sure that the top holes on the shock absorbers are properly aligned with the bottom holes in
the mounting frame (see Figure 19).
30
Installation Guide
Lifting slots