A p p E N D I X B

Connector definitions

Inputs and outputs

NOTICE
You must use copper wires only. Using other types of conductors could damage the
instrument AND void the warranty.
Figure 65 Analytical unit communication ports
1
Connector
1
OpTICAL
Optical Ethernet
2
SErVICE
Ethernet
3
SCADA
Ethernet
4
USB
USB 2.0
CASE
1. CASE
5
2. CASE
2
3
4 5
9
5
Connection
ST-ST full duplex
Category 5
Category 5
A
Shield
6
7
j
Type
Wire
62.5/125 multi-mode
Straight
Straight
n/a
Stranded
AWG #24 to AWG #16
WITh FErrULE
8
5
k
Max. length
1220 m (4003 ft)
3 m (10 ft)
100 m (328 ft)
n/a
n/a
Screwdriver
n/a
n/a
n/a
n/a
Flat Ø 3.5 mm