Run a UV decontamination "extended cycle"

The Magnis/Magnis Dx NGS Prep System includes a UV tube that can be used to decontaminate
the surfaces of the instrument deck. The extended cycle is a 2-hour decontamination. Agilent
recommends running the extended cycle in the event of a spill or leakage on the instrument deck.
The system only permits running the extended cycle once every 7 days to protect the deck
surfaces from excessive exposure to UV light.
1 Verify that the instrument deck is cleared of all labware, then make sure that the instrument
door is closed.
2 From the Home screen, press Decontamination.
The
Figure 25 Decontamination screen with Extended cycle selected
3 At the top of the screen, select Extended cycle, as shown in
4 Press Start.
The decontamination cycle begins and the screen displays a countdown of the time remaining.
At the completion of the cycle, the UV light turns off and the instrument powers down.
If needed, press Abort at any time to stop the decontamination cycle.
Magnis and Magnis Dx NGS Prep System User Guide
Decontamination screen
opens.
Figure
25.
42