Replacing the UV Tube and Viewing UV Tube Usage

Request a UV tube replacement

After 630 hours of UV-tube usage, the next time a decontamination cycle is initiated, the system
notifies you to replace the UV tube. Replacement of the UV tube must be performed by an Agilent
engineer or Agilent authorized service provider. Contact
schedule a UV tube replacement.

View usage hours for the UV tube

After replacing the UV tube, the Agilent engineer or Agilent authorized service provider will reset its
usage tracking to zero. The lifespan of the UV tube is 630 hours.
1 From the Home screen, press Settings.
The
2 Press Hardware Usage Tracking.
3 The
The screen displays the number of hours (h), minutes (m), and seconds (s) of use for the UV
tube.
Figure 32 Hardware Usage Tracking screen
4 Press Close to exit the screen.