6 Maintenance
6.1 Cleaning, disinfection, and sterilization
68
6.1.4.7

High-level disinfection of the window sleeve

NOTE
The hot air sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection (HLD) is as follows –
assuming the CIDEX
your dealer:
A HLD set to support the HLD processes can be ordered from Dentsply
Sirona with REF 66 83 184.
A
HLD tank
B
Cover
1. Carry out a preliminary cleaning process (see "Preliminary cleaning
of the sleeve  [→ 66]").
2. Always use personal protective equipment when implementing the
high-level disinfection if not before.
3. Remove the sleeve from the scanner body (see "Removing the
sleeve  [→ 66]") section.
4. Place the white protective sleeve on the scanner head and place
the scanner body in the scanner holder.
5. Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.
Operating Instructions CEREC Primescan AC, Primescan AC
®
OPA disinfectant is available in your country via
C
Metal bracket
D
Plugs
Dentsply Sirona
66 54 573 D3696
D3696.201.01.05.02    09.2019