Operating Instructions CEREC Omnicam AC / Omnicam AC
66 54 656 D3696
D3696.201.02.03.02    05.2019
6.1.3.8
High-level disinfection of the mirror sleeve (with HLD Set)
NOTE
Dry heat sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection is as follows:
With HLD Set
NOTE
Sporox II should only be used with steel mirror sleeves. Steel mirror
sleeves are marked with the letters ST (see the image).
A HLD set to support the HLD process should be ordered from Dentsply
Sirona with REF 66 05 120.
A
HLD container
B
Lid
1. Pre-cleaning (see „Preliminary cleaning of the mirror sleeve 
[→ 68]").
2. Put on Personal Protective Equipment at the latest when handling
the high-level disinfectant.
3. Remove the sleeve from the CEREC Omnicam / Omnicam (see
„Removing the mirror sleeve  [→ 66]").
NOTE
It is recommended that you never switch between disinfecting and
sterilizing solutions. However if you do, make sure you thoroughly
rinse the HLD set before switching.
4. Place the white protective sleeve on the scanner tip and place the
scanner in the scanner cradle.
5. Use one of the following disinfectants for the high-level disinfection:
®
CIDEX
OPA, Sporox II.
6.1 Cleaning, disinfection and sterilization
C
Metal holder
D
Plug
6 Maintenance
69