Installing/removing and connecting analyzer and option modules

7.4 Installation of analyzer modules

7.4
Installation of analyzer modules
Overview
Figure 7-3
Procedure
1. Isolate the device from power.
2. Remove the housing cover. To do so, unscrew the two cap nuts on the rear panel and push
the housing cover to the back.
3. Remove the blank plate at the rear of the housing at the corresponding slot. To do so, loosen
the four screws and remove the blank plate.
4. Place the analyzer module
5. Fasten the analyzer module
of the housing.
Tighten the screws on the rear panel and the base of the housing with a torque of 2.5 Nm.
58
Rack-mounted housing
OXYMAT 7 analyzer module
Installing an analyzer module, example OXYMAT 7
in the rack-mounted housing
with four screws to the rear panel and two screws to the base
.
Rack-mounted device
Operating Instructions, A5E31874006-12