Instructions for use ESTETICA E70 Vision / E80 Vision
3 Product description | 3.10 Technical data
Operating environment
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
▶ It is essential to comply with the operating conditions specified in the
"Technical Specifications" chapter.
Floor quality
Ambient temperature
Optimum ambient temperature
Relative Humidity
Air pressure
Max. altitude for operation
Maximum loads
Maximum patient weight load
Tray holder of the dentist element -
loadable up to
Assistant element standard tray -
loadable up to
Dentist element - loadable up to
Service table 1568 without locking
system
Service table 1568 with locking sys-
tem

Transportation and storage conditions

Ambient temperature
Relative Humidity
Air pressure
WARNING
The quality of the flooring must meet
the load bearing ability for buildings
DIN 1055 page 3 and have a pres-
sure resistance in accordance with
DIN 18560 T 1.
+10 to +40
15°C to 35°C
30% to 75% non-condensing
700 hPa - 1,060 hPa
up to 3000 m
180 kg
2 kg
1 kg
2 kg
2 kg
5 kg
-20 to +55
5% to 95% non-condensing
700 to 1060 hPa
42 / 178
o
C
o
C / -4 to +131°F