10 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 10.1 Electromagnetic
Measurements of emitted in-
terference
HF emissions according to CISPR
11
HF emissions according to CISPR
11
Emission of harmonics according
to IEC 61000-3-2
Emission of voltage fluctuations/
flicker according to IEC
61000-3-3
Interference im-
munity tests
Electrostatic discharge
(ESD) according to IEC
61000-4-2
10 Information concerning the electromagnetic
compatibility according to EN IEC 60601-1-2

10.1 Electromagnetic Transmissions

This product is designed for use in an environment like the one cited below. The
customer or the user should ensure that the device is used in an environment
of the specified type.
Conformance
Group 1
Class B
Class A
Conforms

10.2 Resistance to electromagnetic interference

This product is designed for use in an environment like the one cited below. The
customer or the user should ensure that the device is used in an environment
of the specified type.
IEC 60601 test levels Compliance level
± 6 kV contact discharge
± 8 kV atmospheric dis-
charge
Instructions for use ESTETICA E70 Vision / E80 Vision
± 2/4/6 kV contact dis-
charge
± 2/4/8 kV atmospheric
discharge
173 / 178
Transmissions
Electromagnetic environment
-
Guidelines
This device uses HF energy for its
internal functions exclusively.
Therefore, the HF emission of the
device is very low and interfer-
ence with adjacent electronic
devices is unlikely.
This product is suitable for use in
all facilities including residential
ones, and facilities that are dir-
ectly connected to a public power
supply that also supplies residen-
tial buildings.
This product is suitable for use in
all facilities including residential
ones, and facilities that are dir-
ectly connected to a public power
supply that also supplies residen-
tial buildings.
This product is suitable for use in
all facilities including residential
ones, and facilities that are dir-
ectly connected to a public power
supply that also supplies residen-
tial buildings.
Electromagnetic en-
vironment -
Guidelines
Floors should be made of
wood or concrete or be
fitted with ceramic tiles.
If the floor is fitted with
synthetic material, the
relative humidity must
be at least 30 %.