Alarms Troubleshooting - ResMed Lumis Series User Manual

Hide thumbs Also See for Lumis Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Problem/possible cause
Wireless coverage may be poor.
The No wireless connection icon
top right of the screen. no wireless network available.
Device may be in Airplane Mode.
Data transfer is not enabled for your device.
My screen and buttons are flashing but there is no alarm sound or message
Software upgrade is in progress.
Displays message: Read only card, please remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position.

Alarms troubleshooting

Problem/possible cause
Display disappears and an alarm is activated
Power failure.
Power cord is disconnected or mains power has been
turned off during therapy.
Displays message: High leak detected, check your water tub, tub seal or side cover
Humidifier may not be inserted properly.
Humidifier seal may not be inserted properly.
Displays message: High leak detected, connect your tubing
Air tubing may not be connected properly.
Mask may be fitted incorrectly.
Displays message: Tubing blocked, check your tubing
Air tubing may be blocked.
16
Solution
Make sure that the device is placed where there is coverage
(ie, on your bedside table, not in a drawer or on the floor).
The Wireless signal strength icon
coverage when all bars are displayed, and poor coverage
when fewer bars are displayed.
Make sure that the device is placed where there is coverage
is displayed on the
(ie, on your bedside table, not in a drawer or on the floor).
If instructed to do so, send the SD card to your care provider.
The SD card also contains your therapy data.
Turn off Airplane Mode, see Travelling by plane.
Talk to your care provider about your settings.
Software upgrade takes approximately 10 minutes to
complete.
Move the switch on the SD Card from the lock position
the unlock position
Solution
Remove your mask until power is restored.
Ensure the power cord is connected and the mains power
switch (if available) is on.
Make sure the humidifier is correctly inserted.
Open the humidifier and make sure that the seal is correctly
inserted.
Make sure the air tubing is firmly connected at both ends.
Make sure your mask is fitted correctly. See your mask user
guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
indicates good
and then re-insert it.
to

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lumis 150 vpap st-aLumis 100 vpap st-a

Table of Contents