12.0 Notes on EMC

Medical electrical equipment is subject to special precautions with regard to EMC and must be installed
acc. to following EMC notes.
Portable and mobile HF communication facilities can influence medical electrical equipment.
The use of other accessories, other converters and cables than stated may lead to an increased emission
or a reduced interference immunity of the equipment or system.
Guidelines and Manufacturer´s Declaration - Emissions
The ATMOS
E 341 Battery is intended for use in the electromagnetic environment specified below. The
®
customer or user of the ATMOS
Emissions Test
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonics IEC 61000-3-2
Flicker IEC 61000-3-3
Guidelines and Manufacturer´s Declaration - Immunity for ATMOS
E 341 Battery
The ATMOS
E 341 Battery is intended for use in the electromagnetic environment specified below. The
®
customer or user of the ATMOS
Immunity Test
IEC 60601- Test Level
ESD IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
EFT IEC 61000-4-4
± 2 kV Mains
± 1 kV I/Os
Surges
± 1 kV common-mode
IEC 61000-4-5
± 2 kV differential
mode
Voltage Dips /
< 5 % U
Dropout
(> 95 % Dip of the U
IEC 61000-4-11
for 0.5 Cycle)
40 % U
(60% Dip of the U
5 Cycles)
70 % U
(30% Dip of the U
25 Cycles)
< 5 % U
(> 95 % Dip of the U
for 5 s)
Power Frequency
3 A/m
50/60 Hz
Magnetic field
IEC 61000-4-8
NOTE
U
is the mains alternating current prior to application of the test levels.
T
E 341 Battery should ensure that it is used in such an environment.
®
Compliance
Group 1
Class B
Class A
Corresponds
E 341 Battery should ensure that it is used in such an environment.
®
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV common-mode
± 2 kV differential
mode
< 5 % U
T
T
(> 95 % Dip of the U
T
for 0.5 Cycle)
40 % U
T
T
for
(60% Dip of the U
T
5 Cycles)
70 % U
T
T
for
(30% Dip of the U
T
25 Cycles)
< 5 % U
T
T
(> 95 % Dip of the U
T
for 5 s)
Inapplicable
Electromagnetic Environment - Guidance
The ATMOS
E 341 Battery uses RF energy
®
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The ATMOS
E 341 Battery is suitable for use
®
in all establishments, including domestic,
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic purposes.
Electromagnetic Environment -
Guidance
Floors should be wood, concrete, or
ceramics tile. If floors are synthetic,
the relative humidity should be at
least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
T
environment. If the user of the
ATMOS
E 341 Battery demands
®
continued function even in case of
for
interruptions of the energy supply,
T
it is recommended to supply the
ATMOS
E 341 Battery from an
®
for
uninterruptible current supply or a
T
battery.
T
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.
®
Notes on EMC
53