3.2 Preparing the device

Prior to fi rst operation peruse the safety notes in chapter „2.0 Hints for your safety" on
page 11.
Damaged pump diaphragms due to cold temperatures during transport.
1. After transport in temperatures below -5°C the device must be acclimatized for up
to 6 hours at room temperature before you continue with the next steps.
2. Check the device for any damage in transport.
3. If the device is damaged: Document and report the transport damage. Send in the
device to ATMOS (chapter „6.3 Sending in the device" on page 40).
4. If the device is not damaged, place it on a safe and even surface.
5. Check the charging accessories for any damage.
6. Damaged charging accessories must be exchanged immediately.
7. The battery must be fully charged, chapter „3.3 Charging the battery" on page 18.
8. Remove the canister system from the support.
9. With DDS canister system: Prior to fi rst use, clean the canister system and insert
a bacterial fi lter (chapter „5.0 Cleaning and disinfection" on page 31 as well as
chapter „3.4 Connection and removal of canister system and hoses" on page 19).
10. Connect the suction hose.
11. Place the canister system upright from above into the support: Chapter „3.4
Connection and removal of canister system and hoses" on page 19.
12. Wrap the suction hose onto the hose rewind.
13. If you wish to use the device for vacuum mattresses: Check whether a suitable
adapter is available for the vacuum mattresses.
Setting up and starting up
17