The pump starts. The vacuum is set which was selected prior to switch off .
»
Patient may be seriously injured, if the set vacuum is too high.
2. Select the vacuum according to the patient and application. Push the button for the
required vacuum.
The green LED above the selected button fl ashes.
»
As long as the auxiliary air vent is opened, the device does not suck.
1. Open the auxiliary air vent before inserting the suction catheter.
2. Apply the suction catheter in such a way as you were taught.
3. Close the auxiliary air vent, so that the device sucks.
Suff ocation is possible by full canister system.
4. Pay attention to the fi lling level of the canister system.
5. Empty the secretion canister or change the Serres
As soon as the canister system is too full, the fl oat ball seals the intake area. You
can then no longer suck with the device.
Make sure that the hose is not kinked during suction because otherwise the suction at
the patient is too low.
If you want to interrupt the suction briefl y, you can clamp the suction hose into the
opening of the hose rewind.
If secretion has penetrated into the device, please observe chapter „5.6 Oversuction" on
page 37.
After use
1. Switch off the device by pressing the on/off button for at least 1 second.
2. Clean the device after every use: Chapter „5.0 Cleaning and disinfection" on page
31.
3. Please carry out a functional check after each cleaning: Chapter „6.2 Function check"
on page 38.