10. Insert the fi lter holder into the canister lid and turn it clockwise until it clicks into
place.
11. Connect the suction hose to the canister system.
12. Place the canister system upright into the support and at the same time position
the suction hose in the hose guide.
13. Check that the clip
canister system.
The vacuum connection from the pump to the canister system is established.
»
14. Check whether the hose has a kink. If applicable, remove the kink.
15. Insert the suction hose into the hose guide of the device base.
16. Wrap the suction hose onto the hose rewind.
17. If required: Connect a fi ngertip to the suction hose.
18. Carry out a manual function check: Chapter „6.2.1 Manual function check" on page
38.
of the canister lid is correctly attached to the support of the
1
1
click
click
Setting up and starting up