5.5 Cleaning and disinfection plan

What
Parts to be
reprocessed
Surfaces
Housing
Device base
Hose rewind
Wall and device support
Power supply and
recharching unit
Canister system
DDS secretion canister
Outer canister lid (DDS)
Inner canister lid (DDS)
Float ball (DDS)
Filter holder
Sealing ring
Bacterial filter
Fingertip
Support for
canister system
Suction hose (DDS)
Recommended disinfectants
Surface disinfection for coated surfaces:
• Green & Clean SK (ATMOS)
• Dismozon
pur (Bode Chemie)
®
• Dismozon
plus (Bode Chemie)
®
• Kohrsolin
FF (Bode Chemie)
®
• Kohrsolin
extra (Bode Chemie)
®
• Perform
(Schülke & Mayr)
®
• Terralin
Protect (Schülke & Mayr)
®
Other surfaces:
• Dismozon
pur (Bode Chemie)
®
• Dismozon
plus (Bode Chemie)
®
• Kohrsolin
FF (Bode Chemie)
®
• Kohrsolin
extra (Bode Chemie)
®
• Mikrobac
forte (Bode Chemie)
®
• Perform
(Schülke & Mayr)
®
• Terralin
Protect (Schülke & Mayr)
®
• SaniCloth
Active (Ecolab)
®
• Incidin
Active (Ecolab)
®
• Mikrozid
Sensitive Wipes (Schülke
®
& Mayr)
2015-07 Index: 01
36
Cleaning and disinfection
Cleaning and disinfection plan
ATMOS
How
C
D
S
Recommendations
X
X
Wipe cleaning and disinfection
X
X
Wipe cleaning and disinfection
1
X
X
Wipe cleaning and disinfection
1
X
X
Wipe cleaning and disinfection
Wipe cleaning and disinfection with a damp
X
X
cloth. Do not immerse into any liquid!
Cleaning with a brush; cleaning and
X
X
X
disinfection (automatic or manual),
1
2
sterilization is possible
Cleaning with a brush; cleaning and
X
X
X
disinfection (automatic or manual),
1
2
sterilization is possible
Cleaning with a brush; cleaning and
X
X
X
disinfection (automatic or manual),
1
2
sterilization is possible
Cleaning with a brush; cleaning and
X
X
X
disinfection (automatic or manual),
1
2
sterilization is possible
Cleaning with a brush; cleaning and
X
X
X
disinfection (automatic or manual),
1
2
sterilization is possible
X
X
Cleaning and disinfection
1
Exchange. If the fi lter is blocked it must also
be exchanged.
Exchange.
X
X
Cleaning and disinfection.
1
X
X
X
When using without suction catheter.
1
2
X
X
X
When using with suction catheter.
1
2
Instruments - manual disinfection:
• Mucocit
-T (Merz Dental)
®
• Gigasept
FF neu (Schülke & Mayr)
®
• Sekusept
PLUS (Ecolab)
®
Please see the manufacturer's
instructions for concentration, contact
time, temperature and the compatibility
of materials.
Wrong concentration of disinfectants
C / E 341 Battery
®
Important information
Wipe cleaning and wipe disinfection: All
surfaces have to be wiped with a clean
(disposable) wipe which is damped with
disinfectant solution; the entire surface
has to be wiped thoroughly and may not
be dried afterwards.
Preferred: machine cleaning and
1)
disinfection in the washer disinfector
(rinse cycle up to 95°C)
If required hot-steam sterilization at
2)
134° C with devices acc. to EN285
* Homecare, unless there is no change
in patient
may lead to damage!
When
Who
Qualifi ed and trained
staff who are familiar
with reprocessing.
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
The above stated hygiene requirements are based
on the regulations according to the Medical Devices
Act, the Medical Devices Operator Ordinance, §18
IfSG and the recommendations of the Robert Koch
Institute. Defi nition of the required reprocessing steps
result from the recommendations of the Robert Koch
Institute: „Requirements for the reprocessing of medical
products". The medical products were categorised in
the risk groups uncritical, semicritical and critical. The
reprocessing measures mentioned in this cleaning and
disinfection plan are a recommendation of ATMOS
MedizinTechnik. Any additional reprocessing measures
are at the operator's discretion. All the recommended
disinfectants which are stated herein are listed
disinfectants (VAH/RKI) and have been tested on their
suitability of use on the ATMOS
MedizinTechnik cannot be hold liable for any damage
caused by wrong concentration of the disinfectants or
by the application of any other disinfectants. Patients
with suspicion of a clinical disease or who developed a
transmissible spongiform encephalopathy (CJK, vCJK,
etc.) have to be treated at facilities which are able to
provide for the necessary preventive measures against
infection. The reprocessing of the reusable instruments
and material may only be performed at facilities which
have an externally certifi ed QM Management acc. to
DIN EN ISO 13485. The Medical Devices Act, IfSG, the
RKI directives, BGR 250 and TRBA 250 always have to be
considered.