4.0 Operation
WARNING
Risk of infection by lack of hygiene or damaged components.
Deadly diseases can be transmitted.
•
Please use new consumables and new disposable canister systems or reprocessed
DDS canister systems for each patient.
•
Prior to each use, please check whether hoses or canister systems are damaged.
Replace any damaged parts.
Electric shock from damaged equipment.
Cardiac arrhythmias may be caused.
•
Prior to each use, please check whether device and charging accessories are
damaged.
•
Replace damaged parts immediately.
•
Do not use the device if it is damaged
Ambient conditions during operation
•
Temperature:
•
Relative air humidity: 5...95 % without condensation
•
Air pressure:
4.1 Switch on the device
The device should only be left on as long as you need it. This way you can increase the
battery life.
1. Push the on /off button to switch on the device.
The pump starts. The vacuum is set which was selected prior to switch off .
»
All LEDs on the control panel light up for about 1 second.
»
The on/off button is illuminated as long as the device is switched on.
»
4.2 Switch off the device
1. Switch off the device by pressing the on/off button for at least 1 second.
4.3 Vacuum adjustment
WARNING
Vacuum is too high.
Patient may be seriously injured.
•
Observe the applicable guidelines.
•
Select the vacuum according to the patient and application.
1. Push the on /off button to switch on the device.
The pump starts. The vacuum is set which was selected prior to switch off .
»
2. Push the button for the required vacuum.
The green LED above the selected button fl ashes.
»
28
Operation
-5...+50° C
540...1100 hPa