1
5. Check whether the foil of the Serres
external canister and the lid tightly fi ts to the Serres
6. Connect the disposable suction hose with the angle
7. Close the auxiliary air vent of the fi ngertip and close the front opening with your
thumb.
8. Switch on the device so that the pump builds up a vacuum.
The Serres
suction bag develops.
»
®
9. Insert the suction hose into the hose guide at the device base.
10. Wrap the suction hose onto the hose rewind.
11. Carry out a manual function check: Chapter „6.2.1 Manual function check" on page
38.

3.5 Support for canister system

3.5.1 DDS canister system

Removal
1. Remove the canister unlocking.
2. Push the support for the DDS canister system backwards up to the middle and take
it out of the guides.
2
suction bag is fully inserted into the Serres
®
external canister.
®
to the Serres
suction bag.
®
2