Emergency
Table of Contents

Advertisement

2. Remove the device base in the following order A - B - C - D:
A
B
Mounting
NOTICE
Incorrectly mounted device base.
Device moves during the operation.
Attach the device base with particular care, according to the operating instructions.
Prerequisite:
The following parts are removed:
Canister system
Support for canister system
Hose rewind
Battery compartment cover
1. Put the device carefully on the front.
2. Take care that the indentations at the device base are fi tted to the protruding edges
at the device.
3. Attach the device base in the following order D - C - B – A. The arrows show the
points at which the device and the device base have to engage with each other.
D
4. Firmly press on all sides again.
5. Afterwards the following parts can be mounted:
-
Battery compartment cover (chapter „6.5 Battery exchange" on page 40)
-
Hose rewind (chapter „3.6 Hose rewind" on page 26)
-
Support for canister system (chapter „3.5 Support for canister system" on page
23)
-
Canister system (chapter „3.4 Connection and removal of canister system and
hoses" on page 19).
D
C
B
A
C
Setting up and starting up
27