5. Hygienic preparation

6. Functional Control

Clean the case inside and out simply by wiping with
suitable disinfectant solution commercially available.
We recommend
Instructions for Use!). You are recommended to
wear suitable gloves (e.g. household or dispos-
able gloves) during disinfection procedures.
Remove the dividers first.
If the ULM CASE III is fitted with oxygen cylinders,
use a clean, dry cloth or a cloth dampened with
clean water to clean the outside of the fittings.
Carry out a functional control after every use(or at
least every 2 years if not used).
If you discover faults during the functional control,
you must not use the emergency case again until the
faults have been rectified.
Carry out the functional control as follows:
1. Check whether the inserts,cover plate, ampoule
strips, corners of the case and the case handle
are free from damage.
2. Check that the case fasteners can be opened
and closed easily.
3. Check the case and cover plate hinges. They
must move easily and be fixed in place correctly.
Replace faulty parts.
®
protect (following the
terralin
Hygienic preparation EN
11