Introduction

Recommended Maintenance

Required Tools and Supplies

Revision L
Maintenance: Introduction
Regular maintenance, irrespective of usage, is essential to ensure that the equipment
will always be functional when required.
WARNING
Failure on the part of all responsible individuals, hospitals or
institutions, employing the use of this device, to implement the
recommended maintenance schedule may cause equipment failure
and possible health hazards. The manufacturer does not in any
manner, assume the responsibility for performing the
recommended maintenance schedule, unless an Equipment
Maintenance Agreement exists. The sole responsibility rests with
the individuals, hospitals, or institutions utilizing the device.
In addition to a standard set of hand tools, you will need the items listed in the table
below.
Item
#10 Torx driver
#6 Torx driver
Phillips screw driver
Leakage current tester
Multifunction micro-simulator
Precision dust remover
Lint-free soft cloth
PS2 style keyboard (Japan only)
MAC™ 3500 Resting ECG Analysis System
2021337-036
Table 1. Tools and Supplies
MT-1216-02AAMI (for 220V)