Passwords

Clinical

Report Format

Editing

Revision L
Troubleshooting: Frequently Asked Questions
Q: Can you set up a password for the Delete function that is different than the
System Setup password?
A: No. The password for the System Setup and the Delete function are the same.
Q: How do I change the way an ECG looks when it prints out?
A: Do the following steps.
Go to System Setup.
Select ECG.
Select which type of ECG report you want to change:
Select unconfirmed reports from the menu.
Find the report type you want the MAC 3500 to print.
Place the number of copies you want in the appropriate column.
If you want the MAC 3500 or 12SL interpretation included on the ECG, put the
number of copies you want in the "with" column.
If you do not want the MAC 3500 interpretation to print on the ECG, put the
number of copies you want in the "without" column.
Click view report type, to see the examples of the report formats.
Press the esc key until you return to System Setup.
Select Save Setup.
Select To System.
You can select Main Menu to exit System Setup.
Q: Can you edit the interpretation at the MAC 3500, and then transmit the edited
record to the MUSE system as an unconfirmed record?
A: If you edit demographic information only the record is still transmitted to the
MUSE system as an unconfirmed record. However, if you edit the
interpretation, the data will not be saved unless the record is confirmed at the
MAC 3500. The record is transmitted to the MUSE system as a confirmed
record as well.
MAC™ 3500 Resting ECG Analysis System
2021337-036
4-31