FRU / Option
Mag Card Reader
Treadmill or Ergometer I/F Cable
Blood Pressure Monitor
Description
No Parts Replaced
Software Update
Hardware Upgrade
Annual Electrical Safety Checkout
Tools
Visual Inspection
5-36
Maintenance: Functional Checkout Procedures
Table 4. Option Repairs (Continued)
Tools
1,2,5
1,2,5
1,2,5
Table 5. Non-FRU Repairs
Tools
1,5
1,5,6
1,2,4,5
1,2,3,5
1.
ECG Simulator
2.
Standard hand tools including a #10 and a #6 Torx Driver
3.
Current Leakage Tester
4.
Anti-static wrist strap
5.
Applicable Service and/or Operator Manual as needed for reference
6.
Floppy Diskette or SD Card
Inspect the following for excess wear and or any visual signs of damage.
1.
Passed check for defective or broken patient cable/leadwires and out of date
electrodes?
2.
Discussed electrode placement, skin prep, and patient related requirements with
ECG Tech?
3.
Keyboard/LCD passed inspection?
4.
AC Power cord passed inspection?
5.
Battery (swollen or leaky) passed inspection?
6.
Trolley and casters passed inspection?
7.
External surfaces passed inspection?
8.
FUSE type and rating passed inspection?
MAC™ 3500 Resting ECG Analysis System
2021337-036
Visual Inspection
Checkout Procedure(s)
9
9
9
Visual Inspection
Checkout Procedure(s)
1,2,3,4,5,6,7
N/A
9,10
7,9,10
1,3,9
1,3,6
1,3
1,3,4
*5 through 9
1,2,3,4
*5 through 9
1,2,3,4,20
*5 through 9
20, 1,2,3
Revision L