Equipment Problems

ECG Data Noise

Revision L
Troubleshooting: Equipment Problems
If the acquired ECG data displays unacceptable noise levels:
When troubleshooting noise or signal quality, be sure the problem is not being
caused by poor skin preparation, or placement and condition of electrodes.
Careful skin preparation is the key to an interference-free ECG. Signal quality is
indicated using Hookup advisor. Click here to learn more about Hook-up
Advisor. Hookup Advisor can be turned on or off in the ECG menu.
Select System Setup  ECG  ECG Acquisition.
Check for defective or date expired electrodes.
Check for defective, broken, or disconnected leadwires.
Run the Acquisition Module Tests in the Diagnostic menu and make sure all
lead wires pass the noise test.