COMM Board Replacement

5-30
Maintenance: COMM Board Replacement
1.
Remove the battery as described in
2.
Using a #10 Torx driver, remove the fasteners from the panel surrounding the
LAN and modem ports.
3.
Grasp the side edges of the COMM board. Rock the board side to side in the
slot as you pull it out along its rails.
4.
Insert the new COMM board as shown below. Slide it onto the rails until it is
seated in place.
MAC™ 3500 Resting ECG Analysis System
2021337-036
"Battery Replacement"
on page 5-6.