Service Manual, Partisol Model 2000 Air Sampler
Figure 3-15. FTS Flow
Calibration screen.
S
3: S
P
ECTION
ERVICE
ROCEDURES
Basic
FTS Flow Calibration
Cur Flow:
Act Flow:
FTS Pres:
Flow Adj: Hold
Offset:
-0.0156
Span:
1.0096
Edit
Hub1
Soft Function Keys in View Data Mode
Edit
Hub1
Calib
Pump
Soft Function Keys in Edit Mode
--
-
+/-
5) While in the Main screen, press <F5: Setup> and then <F2: Calib>
to enter the Calibration screen. From the Calibration screen,
press <F6: FTS> (<SHIFT> <F1>) to enter the FTS Flow Calibration
screen (Figure 3-15).
6) If you are using the Streamline FTS Flow Transfer Standard to
calibrate the flow of the Partisol sampler, enter the FTS calibra-
tion constants "m" and "b" on the FTS Flow Calibration screen.
Execute the following steps to enter these constants:
• Press <F1:Edit>
• Scroll down to the "FTS Const m" field and enter the m constant from the FTS
calibration sheet
• Scroll down to the "FTS Const b" field and enter the b constant from the FTS
calibration sheet
• Press <Enter>
7) Attach the Streamline FTS Flow Transfer Standard to the hub's
sample tube (Figure 3-16). Other flow meters may require the use
of the flow audit adapter to attach the meter to the sample tube.
The flow audit adapter should be installed with its valve open
(Figure 3-8).
8) Do not turn the pump on. If it is currently on, press <F7: Pump>
(<SHIFT> <F2>) to turn it off.
0.1
l/min
0.1
l/min
0.00 in/H2O
FTS Const m:
FTS Const b:
Sat2
Sat3
Sat2
Sat3
DecrFlw HoldFlw
+
++
Revision B.000
Stop
Valve
Sat4
>
Sat4
>
Unshifted
IncrFlw
<
Shifted
Bksp
>
Unshifted
<
Shifted
P
AGE
3-20

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents