Thermo Scientific PARTISOL 2000 Service Manual page 16

Air sampler
Table of Contents

Advertisement

B
Is pump still
Perform system
No
running?
reset.
Yes
Is
Place hand near (do NOT
electronics
touch!) the heater at the
No
compartment
top of the electronics
temperature
compartment.
> 2°C?
Yes
Place multimeter across
test point +24V (red) and
PGND (black) on interface
board.
Check the 4A fuse used
Voltage =~
for the 24.5V@3.6A
No
24V?
voltage from
transformer
Yes
Place multimeter across
test point +6V (red) and
Voltage =~ 6V?
PGND (black) on interface
board.
Yes
Perform system
reset.
Resume normal
Problem
Yes
corrected?
operation.
DISPLAY TROUBLESHOOTING FLOW CHART-B
Resume normal
operation.
Yes
Problem
Go to A
No
corrected?
Heat coming
No
from heater?
terminals and P11 on
the interface board.
Yes
Wait for instrument to
warm up.
Replace fuse
R&P P/N 04-005510-0004
No
Fuse OK?
Check the 5A fuse used
for the 8.3V@5.4A
No
Fuse OK?
voltage from
transformer.
Replace fuse
R&P P/N 04-003681
No
Contact R&P.
Note:
1. Pins for continuity check of cable P16A to P1:
Check continuity
between the heater
Continuity OK?
Yes
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Yes
Replace interface/CPU board
Yes
assembly with known good one
R&P P/N 50-005496 & 10-002407
No
P16A
P1
P16A
P1
Pin 1
Pin 1
Pin 12
Pin 12
Pin 2
Pin 2
Pin 13
Pin 13
Pin 3
Pin 3
Pin 14
Pin 14
Pin 5
Pin 5
Pin 15
Pin 15
Pin 6
Pin 6
Pin 16
Pin 16
Pin 7
Pin 7
Pin 17
Pin 17
Pin 8
Pin 8
Pin 18
Pin 18
Pin 9
Pin 9
Pin 19
Pin 19
Pin 10
Pin 1
Pin 20
Pin 20
Pin 11
Pin 11
Replace heater
No
R&P P/N 56-005909-0120
or 56-005909-0240
Replace transformer
R&P P/N 03-002402
No
Problem
corrected?
Yes
Original boards were
defective.