Summary of Contents for Thermo Scientific Orion Star A113
Page 1
Thermo Scientific Orion Star A113 Benchtop and Star A123 Portable Dissolved Oxygen Meters Reference Guide...
Page 2
Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries. AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc.
Chapter 1 Introduction Thank you for your purchase of the Orion Star A113 benchtop dissolved oxygen (DO) or Star A123 portable dissolved oxygen (DO) meter. These meters are capable of measuring dissolved oxygen levels in % saturation or mg/L units, and temperature of the solution and the membrane.
Connections 1. Power source: a. Power adapter (included with Orion Star A113 benchtop DO meters, sold separately for Star A123 portable DO meters) – Select the appropriate wall socket plug. Slide off the clear plastic cover, and slide on the plug plate into the groove on the back of the adapter.
Primary display Lower, larger display showing measured value in selected mode. Instructions Located below the primary display. These phrases aid in the setup menu and calibration modes. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Meter maintenance can be performed on a daily, weekly or monthly basis, as required by the environment in which the meter is operated. Immediately remove any spilled substance from the meter using the proper cleaning procedure for the type of spill. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Page 10
This page intentionally left blank. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
7.0 Factory Reset a. Press until the desired option is shown. b. Press to save your selection. 4. Press to exit the setup menu and return to measurement mode. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Page 12
The default is no. Before selecting yes, please to the factory defaults and make sure any data that you would like to keep deletes all stored data has been recorded. (calibration and datalog). Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
The oxygen level result from the titration is entered in a manual calibration as the dissolved oxygen value. 4. Press to save configuration and to return measurement mode. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
flash. When the reading has been stable, “READY” will appear. This read type is useful when performing an experiment that requires continuous measurements to be taken, regardless of the measurement stability. Press to save selection. Press the key to return to measurement mode. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Note: Holding the button down will make the value change faster. 4. Press to save configuration and to return measurement mode. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
5. Press to confirm the value. The meter will display slope and will proceed to the measurement mode. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Temperature Measurement and Calibration The Orion Star A113 benchtop dissolved oxygen (DO) and Star A123 portable dissolved oxygen (DO) meter can show the temperatures of the solution or the membrane. To read only temperature and see the temperature on the primary, lower display, follow the instructions below.
Note: The calculated offset will be applied to all future temperature readings. To abort, press to end without saving and return to the measurement mode. 4. When finished, press to save and end calibration. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Chapter 5 Data Storage and Review Orion Star A113 benchtop dissolved oxygen and A123 portable dissolved oxygen meters have a 50 point datalog memory. Automatic Datalog with AUTO-READ Mode ™ 1. In measurement mode, press 2. Press three times in setup until “4.0” is shown on the top line and “READ” is shown on the lower line.
Page 20
This page intentionally left blank. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
If the existing memory’s data can be deleted: 1. Press change to “YES”. 2. Press to clear the datalog. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS Directive are available at www.thermoscientific.com. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Safety requirements for electrical equipment for measurement, CAN/CSA C22.2 No. 61010-1-04 control and laboratory use - general requirements Cheow Kwang Chan Place and Date of Issue: QA/Regulatory Manager June 15, 2011 Singapore Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
0 to 45 ppt Compatible Probe Type Polarographic Temperature Measurement Range 0 to 50°C Resolution Relative Accuracy ±0.1 Offset Calibration 1 point Note: Specifications subject to change without notice. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Value (ppt) This table was calculated from the International Oceanographic Tables, Vol. 1, National Institute of Oceanography of Great Britain, Womley, Godaming, Surrey, England and Unesco, Paris 1971. Thermo Scientific Orion Star A113 Benchtop & Star A123 Portable Dissolved Oxygen Meters...
Need help?
Do you have a question about the Orion Star A113 and is the answer not in the manual?
Questions and answers