Cleaning The Air Intake Filter - Thermo Scientific PARTISOL 2000 Service Manual

Air sampler
Table of Contents

Advertisement

Service Manual, Partisol Model 2000 Air Sampler
Figure 3-6. Location of fan
inside the pump compart-
ment.
S
3: S
P
ECTION
ERVICE
ROCEDURES
5) Place a new fuse with the proper rating into the holder and screw
the two halves of the fuse holder together to secure.
6) Close the electronics compartment door and secure it with the
three screws that you previously removed.
7) Connect the power cord to the AC source and resume normal
operation.
3.1.3.
C
LEANING THE
The Partisol Sampler contains an air intake filter attached to the fan in the pump
compartment. This filter cleans the air flow drawn into the instrument enclosure by the
fan.
Tools Needed: Phillips screwdriver
Materials:
Follow these steps to clean the air intake filter:
1) Turn off the sampler and remove the pump compartment cover
(Figure 3-1).
2) Locate the air intake fan on the left side of the pump compartment
(Figure 3-6).
A
I
F
IR
NTAKE
ILTER
Mild soap solution (if required)
Air intake filter (R&P P/N 10-002799)
Fan
Revision B.000
P
3-5
AGE