Analog Input Calibration - Thermo Scientific PARTISOL 2000 Service Manual

Air sampler
Table of Contents

Advertisement

Service Manual, Partisol Model 2000 Air Sampler
A lways wear appropriate
" 
antistatic devices when
working with the system's
electronics.
Figure 3-12. Main screen.
S
3: S
P
ECTION
ERVICE
ROCEDURES
3.2.2.
A
I
NALOG
NPUT
Please note that the following procedure must be performed after the interface board
calibration and before the temperature, pressure and flow calibrations.
Tools Needed: Multimeter with four-digit resolution
Materials:
Follow these steps to calibrate the analog input:
1) If the Main screen is not currently displayed, press <Esc> until the
Main screen is shown (Figure 3-12).
2) The device must be in the Stop Operating Mode to perform an
analog input calibration. If the hardware is currently in the Run
Operating Mode, press <F4: Run/Stp> to enter the Stop Operating
Mode. The current operating mode is displayed in the upper
right-hand corner of the Main screen.
3) While in the Main screen, press <F5: Setup> and then <F2: Calib>
to access the Calibration screen (Figure 3-13).
4) Plug the 6-pin end of the Analog Input Calibration Cable supplied
with the sampler into the socket labelled "ANEMOMETER" on the
back of the Partisol hub unit (Figure 3-14).
Basic
Partisol Main Screen
16:21 03 - Feb - 00
Stat
Std Vol
H1: DONE
24.143
S2:
ON
16.342
S3:
OFF
S4:
Edit
Stats
Soft Function Keys in View Data Mode
Edit
Stats
Soft Function Keys in Edit Mode
Daily
Today+1
- -
-
C
ALIBRATION
Analog calibration cable (R&P P/N 51-002604)
Curr:2
Valid
24:00
16:21
0.000
0:00
Storage Run/Stp
Storage Run/Stp
+
++
+
++
Revision B.000
Stop
Status:OK
Total
Date
24:00
02-Feb
16:21
03-Feb
0:00
04-Feb
Setup
Setup
Unshifted
Shifted
Bksp
>
Unshifted
Bksp
<
Shifted
P
3-14
AGE