Setting Up - Fuji Electric FWD Series Instruction Manual

Ultrasonic flowmeter for air
Table of Contents

Advertisement

3. Setting Up

It is recommended that the setting of the flow meter be carried out before installation.
For this flow meter, 20 items shown in Table 3-1can be set. On delivery from factory, the "factory
default settings" in Table 3-1 are arranged, so that the flow meter can be used as it is. Change the
setting as necessary to make them suit to your working conditions.
See "4. Display" about button operations and setting methods.
Setting item
F1
Indication and output
Analog output
F2
full scale flow-rate
F3
State of contact output
F4
Contact output
F5
Lower limit alarm flow-rate -59999 to 59999 [m
F6
Upper limit alarm flow-rate -59999 to 59999 [m
Alarm judgment value
F7
hysteresis width
Flow-rate moving average
F8
number of times
F9
Output pulse unit
F10 Pulse output method
Flow-value conversion
F11
selection
Standard conversion
F12
temperature
F13 Test mode time selection
F14 Fluid selection
Current output correlation
F15
value
F16 Low flow cutoff flow-rate
Atmospheric pressure of
F17
the working environment
With or without pressure
F18
value averaging
Reset of all of
F00
accumulated values
FFF
Reset to standard factory
*1
delivery settings
Table 3-1 Setting items and standard factory settings
Forward flow, forward and reverse
flow
0 to 99999 [m
Normal open,
Normal close
Reverse flow pulse, upper/lower
limit flow-rate alarm,
error alarm, electronic statement
output
0 to 9999 [m
01, 02, 04, 08, 16, 32, 64 [times]
Duty or one shot
50, 100, 125, 250, 500 [ms]
Yes (Normal),
Yes (Standard), No conversion
-10 to 60 [°C]
3, 60, Unlimited [minutes]
Air (25A–200A)
Nitrogen (25A–80A)
Instantaneous flow-rate, pressure,
temperature
0 ≤ Set value ≤ Qmin
000.0 to 999.9 [kPa]
With (10 times), without (1 time)
To be cleared, not to be cleared
To be reset, not to be reset
Setting options
3
/h]
3
/h]
3
/h]
3
/h]
*2
3
[m
/h]
12
Factory settings
Forward flow
25A:300
32A:600
40A:700
50A:1200
65A:2000
80A:2500
100A: 5000
150A: 10000
200A: 20000
Normal open
Upper/lower limit flow-rate alarm
3
0 [m
/h]
3
59999 [m
/h]
3
0 [m
/h]
04 [times]
100 [L]
1000 [L] for 100A and larger
Duty
Yes (Normal)
20 [°C]
3
The kind of gas is set as the kind
specified at the time of order.
Instantaneous flow-rate
25A:0.1
32A:0.2
40A:0.2
50A:0.4
65A:0.6
80A:0.8
100A: 2.6
150A: 5.0
200A: 9.0
101.3
With (10 times)
Not to be cleared
Not to be reset