2. Installation

It is recommended that Setting Up (Page12 onward) and changing Panel angle
adjustment: (Page 11) are carried out before installation.
This flowmeter can be installed both indoors and outdoors and on a horizontal pipe and a
vertical pipe.
When a large amount of mist and/or dust exists inside pipe, the vertical piping is
recommended. When installing the flow meter horizontally, install it in such a way that the
display section faces upward.
This flow meter is not of a perfect waterproof structure (IP64). Do not install it at a place where
may be submerged in water.
The use of a sunshade is recommended if the flow meter is exposed to direct sunlight.
If you install the flowmeter in the place where it may get wet with rain, install the flowmeter so
that the display section does not face downward.
When you carry out the piping work, keep hands off inside the body and the ultrasonic
sensors (see Page 6) in particular.
Do not hold the display section of the flow meter.
Flow direction:
Mate the forward flow direction of a fluid to the arrow of the body.