Installation Procedure
7. Refer to
Figure
31. To pair the AIF and DIF together again, set the DIF (2) in position over the
AIF (1). Attach the AIF to the DIF with three screws (3) (0515-0372) included in the kit.
Torque to 9 inch-pounds, starting with the middle screw.
Figure 31
AIF/DIF Pairing
8. Connect the ribbon cable W2 to J820 on the AIF (1).
9. Slide the AIF/DIF assembly into the slot at the rear of the instrument and push on the
assembly to mate the connectors to the motherboard assembly.