Installation Procedure
Figure 11
RF Front End Assembly Removal
5. Using the T-6 driver, remove the four screws (3) securing the A13A1 bandpass filter (4) to the
A13 Front End Assembly. Discard these screws.
6. Disconnect ribbon cable W17 from A13J12.
7. Disconnect A13W1 (5) and A13W2 (6) from A13 Front End Assembly and remove the
combination of A13W1, A13W2, and A13A1 as a unit. This unit will be used as a guide to
assemble the new A13W1 and A13W2 to the new A13A1.
8. Refer to
Figure
12. Locate the spacer, E6607-25005, in the kit. Install the spacer (1) where the
old A13A1 bandpass filter was located. Secure with four 0515-0658 screws (2) and torque to
3 inch-pounds using the T-6 driver.