Installing Microwave Preselector Bypass Switches
1. Refer to
Figure
18. Remove the ribbon cable W18.
Figure 18
Low Band Switch Removal
2. Remove the semi-rigid cables W3, W8, W9, and W10 using the 5/16 inch wrench. These
cables will not be used when the Low Band Switch is re-installed.
3. Remove the two screws (2) using the T-10 driver. The low band switch (1) can now be
removed from the chassis.
Installation Note N9010-90054
Installation Procedure
31