Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following analyzer folder: C:\Program Files\Agilent\licensing.
Verify the License Installation
1. Before the licenses will be recognized, the XSA application must be restarted. Press File, Exit.
An Exit Analyzer dialog box will appear; press Enter to confirm the exit.
2. Double-click on the LaunchXSA icon on the Windows desktop. Wait for the XSA application to
finish starting (the analyzer should be sweeping).
3. Press System, Show, System on the analyzer to display a list of all displayed options. You
should see the following options listed ("x" is 1 for an EXA and 2 for an MXA):
— N90x0A-B40
— N90x0A-B25
— N90x0A-DP2
— N90x0A-MPB
— N90x0A-RTL
— N90x0A-S40
— N9060A-3FP
— N9060A-4FP
— N9060A-5FP
— N9060A-6FP
4. Press System, Show, Errors, Clear Error Queue. This will remove out-of-date errors.
5. Press System, Alignments, Align Now, All.
6. Several alignments will be performed related to optional functionality you just installed.
7. Watch for any errors during the alignments.
8. Press System, Show, Errors to view the instrument error log.
9. Verify that there were no errors during alignment.
Installation Note N9010-90054
Analysis bandwidth, 40 MHz
Analysis bandwidth, 25 MHz
Digital Processor, 2 GB Deep Capture Memory
Microwave Preselector Bypass
Real Time Link
Baseband I/Q Analysis bandwidth, 40 MHz (MXA only)
Amplitude Correction and Limit Lines
TOI and Harmonics
ACP 18 Carrier Measurement
TV Trigger and IQ Analyzer Enhancement
Installation Procedure