Gas analyzer start-up
Checking and approving for use
Performing a final
check of the analyzer
Checking the
environmental
conditions
Checking the
mechanical system
LS4000, LS4060 DIODE LASER ANALYZERS | OI/LS4000/O2-EN REV. A
Topic
Main menu ........................................................................................................... 160
System time menu .............................................................................................. 161
Alignment menu.................................................................................................. 164
Installation procedure ....................................................................................... 165
Ensure that all pre-startup conditions are met.
Proceed in accordance with the following checklists:
Test object
Requirement
Transmitter unit,
The environmental conditions are consistent with the
receiver unit,
following information on the name plate:
junction box
Zone
Explosion group
Temperature class
Ambient temperature
Test object
Requirement
Junction box
The junction box meets the following requirements:
The junction box is not damaged
The junction box has been installed in a stable manner
The junction box is closed so that no dust can penetrate
All of the fixing screws on the housing cover are pre-
sent and securely tightened.