4 MECHANICAL INSTALLATION...

4.1 Locating and Mounting System Items

4.1.1
Model 6553 Gas Monitor – Fig. 4.1
The monitor must be located in the safe area of the application plant in a sheltered interior environment.
The monitor is intended to be panel mounted in a position to suit reading of the displays and with access to the rear to enable
wiring interconnections to be made. The panel preparation requirements and installation dimensions are shown in Fig. 4.1. The
monitor is secured to the panel by two clamping brackets at opposite corners of the monitor chassis.
1 2
3
1 2
3
Kent-Taylor
Purge Gas Monitor
Katharometer
intrinsically
safe circuits
Note. All dimensions nominal
millimeters unless indicated otherwise.
Fig. 4.1 Installation Dimensions and Interconnection Positions – Model 6553 Gas Monitor Unit with Digital Displays
8
290
95.0
%H2 IN AIR
Mounting panel
95.0
thickness range
%H2 IN AIR
entries
Power and signal
conventional
circuits
30
0-20
Cable entry
Cable
3
View on display face of
mounting panel requirements
272
point
278 ± 0.5
3
274
Monitor
case
Area for
clamping
brackets

Advertisement