SECTION 2 – OPERATION
Configure Indicators
This tab allows the operator to change the Name of the input, change the color of the Dial
Indicator on the Home Screen and calibrate the input. Select the appropriate indicator from
the list of indicators on the left. Our service department can perform a calibration procedure
using specialized instrumentation to assure that temperature drift and inaccuracies as a result
of time and usage are compensated for in order to keep your instrument compliant with API
specifications.
Configure Alarms
This tab allows the operator to configure Alarms. When configuring alarms, the following
should be considered:
 Temperature uses Input Signal TC1
 Wall Temperature uses Input Signal TC2
 Pressure uses Input Signal AI1
 Output DO2 controls the Pump (should be disabled on a High-Pressure Alarm)
 Output DO3 controls the Pressure Release
 Output GPO 0 controls the Heater Relay
 Output Relay 1 controls the Heater Contactor (Heater Soft key and Heater switch)
 PID 1 is the Temperature Control Loop
 PID 2 is the Pressure Control Loop