 Schedule Parameters: every schedule must have Schedule Parameters. No segments can
be defined prior to this segment. This segment cannot be deleted.
o End Type: Defines the action to be taken at the End of the Schedule. Options are:
Output Off: The Controller will not be active.
Continuous Hold: The Controller will maintain the Set Point reached at
the end of the schedule.
Reset to Set Point: The End Set Point will become the active Set Point.
End Set Point: The set point to use when the End Type is Reset to Set
Point.
 Segment List: Segments are executed in order. There is a maximum of 13 segments.
o Step Type: Options are:
 Time to Set Point: From the current set point, ramp to the Target Set Point over
the Step Time and then Dwell (maintain the Set Point) for the Dwell Time. Step
Time and Dwell Time can be zero.
 Ramp at Rate: From the current set point, ramp to the Target Set Point at the
Ramp Rate and then Dwell (maintain the Set Point) for the Dwell Time. Dwell
Time can be zero.
 Stop: defines this Segment to be the End Segment. No further segments can be
defined.
 End Step: every schedule must have a defined End Step. No segments can be defined
after this segment. This segment cannot be deleted.
Configure Screen
The Configure Screen allows the operator to customize the Indicators, Buttons and Alarms as
well as Calibrate the inputs.